Description
Zimmer ATS 750 DESCRIPTION
The Zimmer A.T.S. 750 Tourniquet is a full-featured and automatic tourniquet system that is sensitive to the cost issues that all health care professionals face today. Its features include:
- Uses ambient air
- Uses electronic and microprocessor regulated pressure
- Single port system: Uses single port, single bladder cuffs
- 30 minute battery back-up provides mobility by helping ensure continuous power
- Visual and audible alarms activated when an unsafe condition is detected
- Self-check calibration each time the A.T.S. 750 System is switched on
SPECIFICATIONS
General
- Overall Dimensions: 8″ (20.3 cm) H x 7″ (17.8 cm) W x 7.9″ (20.1 cm) D
- Weight: 7.7 lbs. (3.5 kg)
Timer
- Time Alarm Set Ranges: 5 – 240 minutes; 5 minutes increments
- Timer Accuracy: 0.25% of elapsed time
Power
- Line Voltage Range: 100 – 240 VAC, 50/60 Hz., Auto Switching
- Line Current: 670 mA RMS @ 120 VAC
- Input Power: 53 Watts Typical
- Power Cord: Type SJT, AWG 16, 14 ft. (4.27 m)
- Power Plug: Hospital grade, 3 prong straight blade, 15 Amp
Pressure
- Cuff Pressure Range: 50 – 475 mmHg, 5 mmHg increments
- Pressure Accuracy: ±5 mmHg (50 – 475 mmHg)
- Pressure Regulation: ±6 mmHg of set point (10 second average under non-transient conditions without external leaks)
- Maximum Pressures: 475 mmHg cuffs
Battery
- Battery Type: Rechargeable, 12VDC sealed lead acid, 1.2 Amp hours
- Battery Discharge Time: Unit will operate on battery power for 30 minutes minimum with fully charge batteries
- Battery Recharge Time: 24 hours (Maximum) Unit should be plugged in 24 hours before initial use.